• Drg-6002 Thermocouple/Millivolt Signal Alarm Setter
  • Drg-6002 Thermocouple/Millivolt Signal Alarm Setter
  • Drg-6002 Thermocouple/Millivolt Signal Alarm Setter

Drg-6002 Thermocouple/Millivolt Signal Alarm Setter

Certification: RoHS, CCC, ISO: 9001, CE
Condition: New
Customized: Non-Customized
Accuracy: 0.2%F.S.
Input Signal: Thermocouple or Millivolt Signal
Output Signal: 4mA~20madc
Customization:
Manufacturer/Factory & Trading Company

360° Virtual Tour

Diamond Member Since 2024

Suppliers with verified business licenses

Jiangsu, China
Self-branded
The supplier has 2 Self-brands, check the Audit Report for more information
to see all verified strength labels (26)
  • Overview
  • Product Description
  • Detailed Photos
  • Product Parameters
  • Overall dimension
  • Company Profile
Overview

Basic Info.

Model NO.
DRG-6002
Power Supply Mode
20~35V DC
Installation
35mm DIN Rail
Transport Package
Box
Specification
PC+ABSflame retardant
Trademark
PARAGON
Origin
China

Product Description


 
 
Product Description
Drg-6002 Thermocouple/Millivolt Signal Alarm Setter
Detailed Photos
Drg-6002 Thermocouple/Millivolt Signal Alarm Setter
Product Parameters
Basic parameters
Output accuracy(20°C) 0.2%F.S.
Temperature drift 50ppm/K
Response time ≤20ms
Dielectric Strength 1500V AC/1min,2mA
Insulation resistance Above 100M Ω at 500VDC
Electromagnetic compatibility Conform to GB/T18268 (IEC61326-1)
Weight About 150 grams
Drive capability 250VAC/2A or 30VDC/2A
Load capacity Resistance load
Connected on-site equipment Thermocouple
Input data
Input signal Thermocouple or millivolt signal
Input range Thermocouple measurement range
Output data
Output signal 4mA~20mAdc
Output Load 0~500Ω
Power supply
Supply Voltage(Ue) 20V~35VDC
Power protection Power reverse protection
Power consumption About 2W
Power supply mode Independent power supply
Power connection method Terminal power supply, base power supply
Size
Depth 110mm
Height 108mm
Width 18.3mm
Housing
Material PC+ABS flame retardant
Protection grade IP20
Flame retardant grade UL94/V0
Usage environment
Continuous operating temperature '-20ºC~+60ºC
Storage temperature -40ºC~+80ºC
Relative humidity <90% RH (No condensed water)
Environmental requirements There should be no strong vibrations, impacts, high currents, sparks or other electromagnetic induction effects in the surrounding environment. The air should not contain media that corrode chromium, nickel, and silver coatings, and should not contain flammable or explosive substances
Install
Installation method DIN35mm guide rail installation
Overall dimension

Drg-6002 Thermocouple/Millivolt Signal Alarm Setter

 

Company Profile

 


Drg-6002 Thermocouple/Millivolt Signal Alarm Setter
Drg-6002 Thermocouple/Millivolt Signal Alarm Setter
Drg-6002 Thermocouple/Millivolt Signal Alarm Setter
Drg-6002 Thermocouple/Millivolt Signal Alarm Setter
Drg-6002 Thermocouple/Millivolt Signal Alarm Setter
 
Drg-6002 Thermocouple/Millivolt Signal Alarm Setter
Drg-6002 Thermocouple/Millivolt Signal Alarm Setter
Drg-6002 Thermocouple/Millivolt Signal Alarm SetterDrg-6002 Thermocouple/Millivolt Signal Alarm Setter

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now